Fuel testing equipment

CFR IQT-XLM Ignition Quality Tester

Measure ignition delay and derived cetane number (DCN) for diesel fuels in a benchtop constant-volume combustion chamber using ASTM D6890.

CFR IQT-XLM Ignition Quality Tester

Measurement method

The instrument injects the sample into a heated chamber of fixed volume and calculates DCN from the measured ignition delay. Applications in the cut sheet include conventional diesel, biodiesel blends, HVO and GTL.

Methods in the cut sheet
ASTM D6890 · EN 15195 · IP 498
Reported results
Ignition delay in milliseconds and derived cetane number (DCN)
Sample and cycle
30 mL for the test and 35 mL for cleaning. CFR specifies less than 20 minutes per sample.
Power supply
208–240 VAC, 20 A, 50/60 Hz. The built-in UPS supports the cooling system and computer for over 3.5 hours.
CFR IQT-XLM cut sheet · February 2023 (PDF)

Specify your laboratory configuration

Identify the fuel specification, accepted test method and expected sample throughput. Confirm that the laboratory can supply the required compressed air, nitrogen, electrical connection and exhaust ventilation.