Fuel testing equipment

CFR F1/F2 Octane Rating System

Measure the Research Octane Number (RON) and Motor Octane Number (MON) of gasoline on one variable-compression CFR engine.

CFR F1/F2 Octane Rating System

Measurement method

The operator compares knock intensity for the sample and reference fuels. The variable-compression cylinder and four-bowl carburetor support both rating methods. Changing methods involves component changes and settings on the XCP panel.

Research method (RON)
ASTM D2699 · EN ISO 5164 · IP 237
Motor method (MON)
ASTM D2700 · EN ISO 5163 · IP 236
Engine speed
600 rpm for RON and 900 rpm for MON, each ±1%
Data handling
XCP records knock intensity, octane number and test conditions. Excel reporting and LIMS export are supported.
CFR F1/F2 brochure · Form C625, revision B (PDF)

Specify your laboratory configuration

Specify RON, MON or both, and the procedures your laboratory uses. For automatic falling-level tests under Procedure D, specify the optional Octane Analyzer. Identify any intake-air conditioning (EACS) and exhaust requirements.